🎉 BP LabLine Awarded MHEC Contract MC14-D25
Universities and research institutions across New England can now access BP LabLine’s laboratory equipment, research solutions, and scientific consumables through an established MHEC purchasing pathway.
Customize
Want more possibilities ?
Save
$200
Code: RWD200
Ends in 2 Days 23:12:45
Delivery timelines will be provided with your quote
Standard lead time varies by product availability (in-stock items usually 1–3 business days, custom/backordered items 7–21 business days).
90-day return policy
Eligible items may be returned within 90 days after delivery, subject to our Return Policy.
Notice
Final delivery schedules and pricing are subject to confirmation in your official quotation. Customs policies may also impact the total cost. Contact us for your best quote.
overview
Features & Benefits
With a light touch plunger, relaxed finger rest, and an overall user-friendly design, the MicroPette Plus is designed to make pipetting more comfortable and efficient.
Covers 0.5 µL to 1 mL; available in single-channel adjustable, single-channel fixed, and multi-channel adjustable models for exceptional versatility across laboratory applications.
Compatible with standard 96-well plates, making the MicroPette Plus ideal for high-throughput applications.
The easy-to-use rotating counter knob and digital display allow for quick and accurate volume setting, with the selected volume clearly visible at a glance.
Made of heat-resistant materials that allow autoclaving or UV disinfection without disassembly. The design is also resistant to harsh chemicals.
The one-touch tip ejector makes switching between different pipette tips fast and effortless.
Advantages
Specifications
| Parameter | 7030303004 | 7030303006 | 7030303012 |
|---|---|---|---|
| Basic Specification | |||
| Volume Range | 0.5–10 µL | 5–50 µL | 50–300 µL |
| Increment | 0.1 µL | 0.5 µL | 5 µL |
| Channels | 8 | ||
| Test Volume | 10 µL | 5 µL | 1 µL | 50 µL | 25 µL | 5 µL | 300 µL | 150 µL | 50 µL |
| Systematic Error & Random Error | |||
| Test Volume | 10 µL | 50 µL | 300 µL |
| Systematic Error | ±0.15 µL / ±1.50% | ±0.50 µL / ±1.00% | ±2.10 µL / ±0.70% |
| Random Error | ±0.15 µL / ±1.50% | ±0.25 µL / ±0.50% | ±0.75 µL / ±0.25% |
| Test Volume | 5 µL | 25 µL | 150 µL |
| Systematic Error | ±0.125 µL / ±2.50% | ±0.375 µL / ±1.50% | ±1.50 µL / ±1.00% |
| Random Error | ±0.125 µL / ±2.50% | ±0.25 µL / ±1.00% | ±0.75 µL / ±0.50% |
| Test Volume | 1 µL | 5 µL | 50 µL |
| Systematic Error | ±0.04 µL / ±4.00% | ±0.15 µL / ±3.00% | ±0.75 µL / ±1.50% |
| Random Error | ±0.04 µL / ±4.00% | ±0.10 µL / ±2.00% | ±0.40 µL / ±0.80% |